The device was returned to zoll medical united kingdom for evaluation and the device performed to specification.A review of the device log shows there were two "shock advised" events.Both analysis events recorded cpr waveform activity throughout the analysis period which introduced significant artifact into the ecg.The presence of this artifact obscured the underlying rhythm and shifted the analysis results to a "shock advised" result.The device charged on both prompts; however, they were not delivered as aed mode was manually exited which disarms the charged device.All analyses during the case started at the correct interval and the device recorded advisory prompts.The device was put through extensive functional testing including shock functionalities without duplicating any fault to the device.Bench handling with the returned multifunction cable observed noise on the ecg, the multifunction cable was scrapped as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|