BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G447 |
Device Problems
Failure to Capture (1081); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/19/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).It was noted that a follow-up appointment has been scheduled for the patient to come into clinic.No adverse patient effects were reported.Currently, this crt-d system remains in service.
|
|
Event Description
|
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).It was noted that a follow-up appointment has been scheduled for the patient to come into clinic.At follow-up, the left ventricular (lv) lead capture threshold was high at 5.5v so it was reprogrammed to a new vector.No adverse patient effects were reported.Currently, this crt-d system remains in service.
|
|