This is one of twenty manufacturer reports being submitted for this case.Please reference related manufacturer report no: 2015691-2024-02917, 2015691-2024-02918, 2015691-2024-02920, 2015691-2024-02921, 2015691-2024-02930, 2015691-2024-02931, 2015691-2024-02955, 2015691-2024-02960, 2015691-2024-02968, 2015691-2024-02970, 2015691-2024-02971, 2015691-2024-02974, 2015691-2024-02978, 2015691-2024-02980, 2015691-2024-02982, 2015691-2024-02983, 2015691-2024-02985, 2015691-2024-02987, 2015691-2024-02989.Citation: nidal jammoul, md, valentin dupasquier, md, mariama akodad, md, phd, pierre-alain meunier, md, lionel moulis, md, sonia soltani, jean-christophe macia, md, pierre robert, md, laurent schmutz, md, matthieu steinecker, md, christophe piot, md, frederic targosz, md, henri benkemoun, md , benoit lattuca, md, phd, francois roubille, md, phd, guillaume cayla, md, phd, and florence leclercq, md, phd montpellier, france; chu montpellier, france; nimes, france.''long-term follow-up of balloon-expandable valves according to the implantation strategy: insight from the directavi trial'' per the instructions for use (ifu), conduction system defects (heart block), arrhythmias and conduction system defects that may require a permanent pacemaker are potential adverse events associated with balloon aortic valvuloplasty, bioprosthetic heart valves, and the thv procedure.According to the valve academic research consortium (varc) guidelines, the close anatomical relationship between the aortic valve complex and the branching atrioventricular bundle may explain these complications of the thv procedure.According to the literature review, and as documented in ew clinical technical summary for complaints-conduction disturbances/ heart block, atrioventricular conduction disturbances after thv are associated with many patient-related and procedural related factors, including pre-operative co-morbid status, the degree, and bulkiness of aortic valve and annular calcification, inter-ventricular septal thickness, pre-existing electrocardiogram abnormalities, the depth of prosthesis implantation, and the profile of the implanted prosthesis.Unlike conventional avr, where there may be localized trauma due to decalcification of the annulus and/or suture placement in the proximity of the av node or the bundles, thv may cause conduction abnormalities through mechanical impingement of the conduction system by the prosthesis.The mechanisms of the development of heart block after thv are well documented and described in the literature.It is also documented that pre-existing heart block is common in patients undergoing thv or surgical avr and another 4-6 % will develop postoperative heart block, potentially requiring a permanent pacemaker.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate a definitive root cause could not be determined due to limited information received.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time at this time.
|