Model Number G124 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this system with a cardiac resynchronization therapy defibrillator (crt-d) and a left ventricular (lv) lead showed early lv over sensing of myocardial signals that led to bi ventricular pacing inhibition.The crt-d and the lv lead remain in service.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that this system with a cardiac resynchronization therapy defibrillator (crt-d) and a left ventricular (lv) lead showed early lv over sensing of myocardial signals that led to bi ventricular pacing inhibition.Additional information was received mentioning that the health care professional made a programming change related to sensing coils and the system showed no oversensing.The crt-d and the lv lead remain in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|