It was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4, a flail, and a prolapsed posterior.An xtw clip was inserted and both leaflets were able to be grasped.However, while removing the lock line (ll), it got stuck.Troubleshooting was performed, but the ll was still unable to be removed.The physician decided to use a snare as a precaution when removing the clip.The clip was unable to pass through the tip of the steerable guide catheter (sgc).Therefore, the clip was brought to the femoral region and a surgical procedure was performed to removed the clip.The sgc was prepped again to be inserted on the left side of the patient.Upon inserting the dilator during preparation, a loss of fluid column occurred.Therefore, the sgc was replaced.Two clips were then implanted, reducing mr to a grade of 2.There were no clinically significant delay in the procedure.
|