The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection, and magnetic sensor functionality test of the returned device were performed following bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and the device was recognized and visualized on the system, however error 106 appeared on the system due to an open circuit on the tip area.A manufacturing record evaluation was performed for the finished device (b)(4) number, and no internal action related to the reported complaint condition were identified.The force sensor error could be related to the visualization issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle as part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
It was reported that the patient underwent a intrauterine fetal supraventricular tachycardia (isvt) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the medical team identified that the ablation catheter would jump around on the carto 3 system screen when they were ablating.The metal values for the mapper were high when on ablation and were normal when not ablating.The medical team verified that the grounding pads were an appropriate distance from the back patches.They raised the ii and the issue remained.Afterward, they replaced the cable without resolution.The physician wanted to continue the procedure so the catheter was not replaced.No patient consequences were reported.
|