Model Number CI-1600-05 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/20/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient reportedly experienced decreased performance.A review of the recipient¿s test data indicated impedance issues.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top and bottom cover of the device.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Search Alerts/Recalls
|