Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2024
Event Type  malfunction  
Manufacturer Narrative
H10: g: phone: (b)(6).
 
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had an autozero failure.There was no patient involvement when the issue occurred.No patient or user harm reported.A philips remote service engineer (rse) noted that the customer's v60 ventilator had an autozero failure.The customer reported that the alarm listed solenoids 1 and 4 were not detected and had buttons for solenoid 1 and 4 greyed out.The customer reported that the device had recently had a flow sensor assembly replaced.The rse advised the customer to check the pins on the solenoids and ensure that the pins were seated correctly with the data acquisition board.The rse recommended that the customer replace the solenoids and/or data acquisition board.The customer was provided with the part number for a replacement.Investigation ongoing.
 
Manufacturer Narrative
Insufficient information is available to determine the resolution of the event.Multiple good faith efforts (gfe) were performed for further details regarding the event; however, no response was provided.No further details could be obtained regarding the event, and it could not be determined if the customer's issue was resolved or if the device had been repaired.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19154686
MDR Text Key340860198
Report Number2518422-2024-21351
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/01/2024
Date Device Manufactured07/10/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-