It was reported that this implantable cardioverter defibrillator (icd) had a magnet placed on it for a procedure.It was noted that the device delivered shocks during the procedure.Additionally, there was a delayed charge time that resulted in the device reaching elective replacement indicator (eri).The device was able to be taken out of eri.Technical services (ts) performed a data analysis and stated that the delayed charge times were due to intermittent magnet sensing and exposure to the magnetic field.The analysis revealed the battery status is normal, but the programmer latched onto the date eri was declared.Ts stated that it may be difficult to monitor battery status as a result, but the device will produce tones when eri is declared again.The device remains in use.No adverse patient effects were reported.
|