Model Number G179
Device Problems Premature Discharge of Battery (1057); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2024
Event Type  malfunction  
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.The patient reported hearing beeping tones a couple of weeks ago, however, did not have a specific date.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Device replacement was recommended, and a replacement interval was discussed.No adverse patient effects were reported.This device remains in service.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19155023
MDR Text Key340799609
Report Number2124215-2024-24231
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/21/2020
Device Model NumberG179
Device Catalogue NumberG179
Device Lot Number189011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-