Catalog Number 3504001BC
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Capsular Contracture (1761); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/30/2023
Event Type  Injury  
Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event Description
It was reported that a patient implanted with two 400cc mentor memorygel breast implants experienced bilateral capsular contracture and a rupture on the left side post-operatively.In addition, symptoms including pain, hair loss, fatigue, vision changes, and poor memory.Were reported.As a result, the patient underwent explantation on (b)(6) 2024.This report is for the right side device.
 
Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation. mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the sm mpp gel 400cc returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.  at the present time, there is no sufficient evidence to show an association between breast implants and generalized illness (assessing the risks of breast implants and fda¿s vision for the national breast implant registry).As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Type of Device
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19157436
MDR Text Key340758427
Report Number1645337-2024-04755
Device Sequence Number1
Product Code FTR
UDI-Device Identifier00081317000327
UDI-Public00081317000327
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030053
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Catalogue Number3504001BC
Device Lot Number6493553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2024
Date Manufacturer Received03/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/26/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient EthnicityHispanic
-
-