Model Number HL-90
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.Device evaluation: one device was returned for investigation.Visual inspection found: water tank cover gasket is worn out, water tank cover cracked, and a non-oem line cord.The device also had out dated quick connect, pcb, and power switch.Functional testing also confirmed the complaint.The water tank cover was cracked around the screw hole on the bottom right.What caused the crack was not established.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Event Description
It was stated that the device is cracked on the back housing.Issue found during testing.There was no patient involvement and no harm/adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key19159596
MDR Text Key341253994
Report Number3012307300-2024-02903
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(11)021114
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation 505
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2024
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-