U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72081-01
Device Problem High Readings (2459)
Patient Problems Hypoglycemia (1912); Dizziness (2194); Shaking/Tremors (2515)
Event Date 03/28/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and sensor kits were reviewed and the dhrs showed the libre sensor and sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A high readings issue with the adc device was reported.The customer received unspecified higher sensor scans and self-treated with insulin (dose/type unknown), but experienced hypoglycemic symptoms such as dizziness and tremors.An ambulance was called upon arrival, blood glucose levels were measured and the customer was administered with glucose (iv, dose unspecified) and was asked to eat and drink something.There was no report of death or permanent impairment associated with this event.
 
Event Description
A high readings issue with the adc device was reported.The customer received unspecified higher sensor scans and self-treated with insulin (dose/type unknown), but experienced hypoglycemic symptoms such as dizziness and tremors.An ambulance was called upon arrival, blood glucose levels were measured and the customer was administered with glucose (iv, dose unspecified) and was asked to eat and drink something.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection performed on the returned sensor patch and no issue was observed.Data was extracted using approved software and extraction was successful.The watermark was observed at the base of the tail indicating the sensor being properly inserted.Sensor state 7 with event logs 11 and sensor state 8 with event log 9 are an indication that the sensor had terminated due to an error recognized by the sensor.This is a part of software design and is not an indication of product non-conformance.Smu (source measurement unit) test will be performed on the sensor to verify if sensor is functioning as intended.The returned sensor was further investigated and de-cased.Performed an internal visual inspection on the sensor¿s pcba (printed circuit board assembly); no issues were observed.Performed an smu (source measurement unit) test using connector key to ensure the sensor's electronics were functioning correctly and the returned unit did not have any glucose reading issues.Poise voltage and sensor thermistor testing were both within specification, indicating the sensor was providing accurate glucose readings.The passing of functionality (smu) testing is an indication that there were no issues with sensor functionality and electronics.Therefore, this issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19159961
MDR Text Key340791435
Report Number2954323-2024-13744
Device Sequence Number1
Product Code QLG
UDI-Device Identifier35021791002659
UDI-Public(01)35021791002659(17)2025-01-31(10)T60001718(91)72114-01
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K213996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number72081-01
Was Device Available for Evaluation? No
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-