• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Hypoglycemia (1912); Tachycardia (2095); Shaking/Tremors (2515)
Event Date 04/06/2024
Event Type  Injury  
Event Description
An error message was reported with the abbott diabetes care (adc) device.The customer received unspecified sensor error message and was unable to obtain glucose readings.The customer experienced symptoms of shaking hands and heart palpitations.The customer was unable to treat themselves due to symptoms and received an unspecified sweet drink from a non-healthcare professional (customer's husband).There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed on the sensor patch.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) and poise voltage testing were performed and all results were within specification.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An error message was reported with the abbott diabetes care (adc) device.The customer received unspecified sensor error message and was unable to obtain glucose readings.The customer experienced symptoms of shaking hands and heart palpitations.The customer was unable to treat themselves due to symptoms and received an unspecified sweet drink from a non-healthcare professional (customer's husband).There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19160726
MDR Text Key340789788
Report Number2954323-2024-13748
Device Sequence Number1
Product Code QLG
UDI-Device Identifier05021791002504
UDI-Public(01)05021791002504(17)2025-05-31(10)KTP008139(91)71990-01
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/22/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-