MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRON
|
Back to Search Results |
|
Model Number GALI 4LV SONR CRT-D 2844 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/27/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
Reportedly, before implantation, the device was tested and the charge time was > 15 sec and the battery voltage was 2.88v.The device has not been implanted.
|
|
Search Alerts/Recalls
|
|
|