• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE LEAD Back to Search Results
Model Number 0672
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Device Dislodged or Dislocated (2923)
Patient Problems Shock from Patient Lead(s) (3162); Twiddlers Syndrome (4563)
Event Date 04/08/2024
Event Type  Injury  
Event Description
It was reported that this right ventricular (rv) lead delivered inappropriate shocks due to oversensing.A chest x-ray revealed that the lead had dislodged, and the atrial lead had no slack.The dislodgement was a result of twiddler's syndrome.The lead was explanted and replaced.The atrial lead remains in use as the slack returned after the leads were untangled.No additional adverse patient effects were reported.The lead was retained by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19164737
MDR Text Key340838581
Report Number2124215-2024-24558
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526592829
UDI-Public00802526592829
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0672
Device Catalogue Number0672
Device Lot Number234807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-