Manufacturer's investigation conclusion: a direct correlation between heartmate 3 left ventricular assist system (lvas), serial number (b)(6), and the reported stroke could not be conclusively established through this evaluation.Additionally, the reported low flow alarms were confirmed through review of the submitted log files.A specific cause for the reported alarms could not be conclusively determined through this evaluation.Review of the submitted log files revealed multiple low flow alarms.The low flow events appeared to be associated with elevated pulsatility index values.These occurred while the pump was operating at a fixed speed of 4800 revolutions per minute (rpm), with a low speed limit (lsl) of 4600 rpm.The low flow events appeared to be associated with elevated pi values.Following a speed increase to 5000 rpm and then to 5200 rpm, no further low flow alarms occurred.The pump appeared to function as intended at the set speed.Multiple attempts for additional information were sent to the customer; however, no further information has been provided at this time.The patient remains ongoing on heartmate 3 lvas, serial number (b)(6).No product available for evaluation.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviation from manufacturing or quality assurance specifications.The heartmate 3 left ventricle assist system (lvas) instructions for use (ifu) is currently available.Section 1, "introduction," lists stroke as a potential adverse event that may be associated with the use of heartmate 3 lvas.Section 4, ¿system monitor,¿ explains that the low flow hazard alarm occurs when pump flow is less than 2.5 liters per minute (lpm).A 10-second delay is imposed between the detection of the low flow status and the activation of the associated audio and visual indicators on the system controller.Changes in patient conditions can result in low flow.Section 6, ¿patient care and management¿ (under ¿anticoagulation¿), provides the recommended anticoagulation regimen, including international normalized ratio (inr) range, as well as suggested anticoagulation modifications.Section 7, ¿alarms and troubleshooting,¿ explains all system alarms and the recommended actions associated with them.The heartmate 3 lvas patient handbook also outlines all system controller alarms as well as how to respond to each alarm condition.No further information provided.The manufacturer is closing the file on this event.
|