Model Number 27074
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Event Description
During resmed evaluation of an astral device, review of the device data logs revealed total power failure alarms.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to an intermittent battery signal.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 DBL KIT FRA
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED PTY LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key19168642
MDR Text Key340876477
Report Number3007573469-2024-00100
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270743
UDI-Public(01)00619498270743(10)1391881
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number27074
Device Catalogue Number27074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/23/2024
Distributor Facility Aware Date03/25/2024
Device Age5 MO
Date Report to Manufacturer04/23/2024
Date Manufacturer Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-