• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMP SWVLK C,CLD 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMP SWVLK C,CLD 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number BIO-COMP SWVLK C,CLD 4.75X19.1MM
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  malfunction  
Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Event Description
It was reported that during a rotator cuff repair surgery the anchor had no hold after correct insertion.Complete removal was possible.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP SWVLK C,CLD 4.75X19.1MM
Type of Device
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19174654
MDR Text Key341071798
Report Number1220246-2024-02328
Device Sequence Number1
Product Code MAI
UDI-Device Identifier00888867097988
UDI-Public00888867097988
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K203495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBIO-COMP SWVLK C,CLD 4.75X19.1MM
Device Catalogue NumberAR-2324BCC-1
Device Lot Number14463986
Was Device Available for Evaluation? No
Date Manufacturer Received06/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-