ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Hypoglycemia (1912); Convulsion/Seizure (4406)
Event Date 04/11/2024
Event Type  Injury  
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection has been performed on sensor and no issues were observed.The sensor plug was seated in mount properly.Extracted data from returned sensor using approved software.Removed the sensor plug and inspected the plug assembly, no failure mode observed.Sensor was reprogrammed and simvivo test (simulation of the electrical signal produced by the sensor tail) performed.All results were within specification.No malfunction or product deficiency was identified.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.All pertinent information available to abbott diabetes care (adc) has been submitted.
 
Event Description
A "sensor error" message was reported with the abbott diabetes care (adc) device and the customer was unable to obtain readings.As a result, the customer experienced shaking and had a seizure.The customer was unable to self-treat, requiring treatment with "glucogene' and juice and crackers provided by a non-healthcare professional (non-hcp).There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19175202
MDR Text Key340955427
Report Number2954323-2024-14081
Device Sequence Number1
Product Code QLG
UDI-Device Identifier05021791001033
UDI-Public(01)05021791001033(17)250531(10)240113Q(91)71994-01
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2024
Initial Date Manufacturer Received 04/12/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-