Model Number 3851
Device Problem Failure to Advance (2524)
Patient Problems Hemorrhage/Bleeding (1888); Perforation (2001); Vascular Dissection (3160)
Event Date 04/10/2024
Event Type  Death  
Event Description
It was reported that the patient died.The 90% stenosed target lesion was located in the mildly tortuous left anterior descending artery (lad).A 10mmx2.50mm wolverine coronary cutting balloon and 5.50mm x 12mm nc emerge were selected for use.During the procedure, a non-boston scientific ivus catheter advanced to view the proximal to middle left anterior descending artery, however, it was unable to cross the lesion.Then a wolverine was used but it was also unable to cross the lesion.While attempting to cross the lesion several times, dissection and perforation occurred in the left main trunk to the lad proximal with the guide wire, so the nc emerge balloon was used to treat it, but it could not stop the bleeding.The patient died.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19175707
MDR Text Key340964416
Report Number2124215-2024-24701
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0032051404
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IVUS CATHETER - PHILIPS
Patient Outcome(s) Other; Death;
-
-