• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV WITH SHUNTASSISTANT 25; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV WITH SHUNTASSISTANT 25; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV414T
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Hydrocephalus (3272)
Event Date 04/04/2024
Event Type  Injury  
Event Description
It was reported that a progav shunt system (#fv414t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the shunt system showed an over-drainage and adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2024.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 4 years, 7 months.Weight: 19 kgs.Height: 108 cm.Gender: female.
 
Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damage of the valve were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves do not operate within the permissible tolerance in their respective relevant positions.An accelerated outflow of both valves could be determined.Adjustment test: the progav was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in both valves.To make the deposits in the shunt system more visible, they were colored using a staining solution.Results: based on our investigation results, we can determine an accelerated outflow in the horizontal position on the progav, and the shuntassistant also exhibits a slowed outflow in the horizontal position and an accelerated outflow in the vertical position.The deposits visible in the valves may have led to the change in the flow rate.Deposits caused by natural substances in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV WITH SHUNTASSISTANT 25
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key19175979
MDR Text Key340967920
Report Number3004721439-2024-00099
Device Sequence Number1
Product Code JXG
UDI-Device Identifier04041906132118
UDI-Public4041906132118
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K103003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model NumberFV414T
Device Catalogue NumberFV414T
Device Lot Number20059542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age4 YR
Patient SexFemale
Patient Weight19 KG
-
-