• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH Back to Search Results
Model Number CDVRA500Q
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/02/2024
Event Type  malfunction  
Event Description
It was reported that the patient exhibited difficulty in pairing their implantable cardioverter defibrillator (icd) to the merlin app.Upon further investigation, it was found that the icd exhibited loss of bluetooth telemetry functionality.The patient was brought into clinic and the device bluetooth was successfully reset, resulting in successful pairing.There were no patient consequences.
 
Manufacturer Narrative
The reported event of inability to communicate via ble was confirmed.Based on the review of the information provided, it was confirmed that a memory corruption had occurred, resulting in the loss of ble connectivity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19176003
MDR Text Key340974453
Report Number2017865-2024-39535
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public(01)05415067031990(10)P000123586(17)230531
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCDVRA500Q
Device Lot NumberP000123586
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/05/2024
Initial Date FDA Received04/24/2024
Supplement Dates Manufacturer Received04/29/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RIGHT VENTRICULAR (RV) LEAD
Patient Age79 YR
Patient SexMale
-
-