The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored and verified, during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time, product has not yet been returned.And a valid serial number has not been provided.An extended investigation has been performed for the reported complaint.And there was no indication, that the product did not meet specification.Clinical data was reviewed and confirmed, that freestyle libre sensors continue to be safe, effective and perform as intended in the field.Stability data for freestyle libre sensors was reviewed.And showed no anomalies or non-conformances that could have led to the complaint.A tripped trend review was conducted for the reported complaint and freestyle libre sensors.No trends were identified, that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date the incident occurred is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
An email complaint was received, about a high readings issue with the adc device.A customer received, unspecified higher sensor scan results and experienced feeling "appartic".The customer¿s, ¿blood test¿ result was obtained (¿below 50mg¿) and was provided dextro-energy and coca-cola for treatment by a non-healthcare professional.It was indicated, that emergency services were called.And the customer¿s blood glucose level was checked (result unspecified).There was no report of death or permanent impairment associated with this event.
|