• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problems Failure to Interrogate (1332); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Event Description
It was reported that the patient exhibited difficulty in pairing their implantable cardioverter defibrillator (icd) to the merlin app.Upon further investigation, it was found that the icd exhibited loss of bluetooth telemetry functionality.No changes were reported.There were no patient consequences.
 
Event Description
New information received notes the patient was brought into clinic and the device bluetooth was successfully paired via interrogation.There were no patient consequences.
 
Manufacturer Narrative
The reported event of ble lockout was confirmed.Based on the information provided, the ble lockout occurred due to insufficient authorization.It was unable to be determined what triggered the insufficient authorization.This is per device behavior as part of the cybersecurity feature; there is no malfunction suspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19177561
MDR Text Key341142510
Report Number2017865-2024-39575
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public(01)05415067032010(10)S000078384(17)220930
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model NumberCDHFA500Q
Device Lot NumberS000078384
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/19/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATRIAL LEAD; LEFT VENTRICULAR LEAD; RIGHT VENTRICULAR (RV) LEAD
Patient Age64 YR
Patient SexFemale
-
-