• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE Back to Search Results
Catalog Number 50000000E
Device Problems Decreased Pump Speed (1500); Failure to Calibrate (2440)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2024
Event Type  malfunction  
Event Description
It was reported that arctic sun device failed calibration error 2 (pre-warm inlet pressure error).Expected -7.0 and actual 0.06.It was determined during troubleshooting that the circulation pump has failed.They would replace the pump in house.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Manufacturer Narrative
The reported issue was confirmed.The root cause identified as a circulation pump failure.Expected -7.0 and actual 0.06.It was determined during troubleshooting that the circulation pump has failed.A dhr is not required as the reported event is not an out of box failure and therefore the reported event is not manufacturing related.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that arctic sun device failed calibration error 2 (pre-warm inlet pressure error).Expected -7.0 and actual 0.06.It was determined during troubleshooting that the circulation pump has failed.They would replace the pump in house.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19180175
MDR Text Key341580089
Report Number1018233-2024-02279
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-