• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SYNERGYRF¿ CONSOLE; ELECTROSURG CUT/COAG DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SYNERGYRF¿ CONSOLE; ELECTROSURG CUT/COAG DEVICE Back to Search Results
Model Number SYNERGYRF¿ CONSOLE
Device Problem Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The assembly has been evaluated and found that the rf probe connector was damaged.The condition of the device is consistent with questionable user handling.The rf probe connector was replaced.
 
Event Description
It was reported the front port on ar-9800, console the port is melted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGYRF¿ CONSOLE
Type of Device
ELECTROSURG CUT/COAG DEVICE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19180983
MDR Text Key341791172
Report Number1220246-2024-02376
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00888867142473
UDI-Public00888867142473
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161581
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSYNERGYRF¿ CONSOLE
Device Catalogue NumberAR-9800
Device Lot Number11236812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2021
Date Manufacturer Received07/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-