This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field h3, h4 and h6.Corrected fields: d4 and d9.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over four (4) years since the subject device was manufactured.Based on the results of the investigation, the cause for the reported issue of the cracked ceramic tip cannot be definitely determined.However, the most likely causes are: thermo-mechanical fatigue.Wear and tear.Improper handling (impact, shock).Before using the product, the warning notices in the instructions for use should be followed to ensure a faultless condition of the product.See especially chapter 4: ¿warning: infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can.Cause infection of the patient and/or medical personnel.4.1 inspection and testing.Inspecting the product: visually inspect the product.Make sure that it has: no corrosion, no dents, no scratches.Ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury: impact, fall, shock, or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿.Olympus will continue to monitor field performance for this device.
|