BAXTER HEALTHCARE CORPORATION; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number ASKU
Device Problems Loose or Intermittent Connection (1371); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2024
Event Type  malfunction  
Manufacturer Narrative
G1 manufacturing facility - this device was manufactured at one of the two following manufacturing sites: baxter healthcare - mountain home 1900 n highway 201 mountain home, ar 72653 united states.Baxter healthcare - dominican republic carretera sanchez km 18.5, parque industrial itabo, piisa haina, san cristobal 91000 dominican republic.An alarm indicative of a potential malfunction of the disposable cassette was reported.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.However, a loose connection was reported between the supply line of the homechoice cassette and the supply bag, which is known to cause this alarm.The cause of the loose connection could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice (hc) device experienced a system error 2240 (air in line/set) alarm.The patient was connected at the time of the alarm.This occurred during dwell one of four of peritoneal dialysis therapy.During troubleshooting, it was reported that there was a loose connection between the supply line with the blue clamp of the homechoice cassette and the supply bag that led to this alarm.Renal therapy services (rts) assisted the patient with clearing the alarm and in removing the supplies.Proper procedures per the user manual were reviewed with the patient.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE CORPORATION
see h10
see h10
see h10
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19183018
MDR Text Key341058553
Report Number1416980-2024-01899
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/30/2024
Initial Date FDA Received04/25/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HOMECHOICE (HC) CLARIA DEVICE; UNSPECIFIED PD SOLUTION
-
-