It was reported that during a cryo ablation case, a pericardial effusion was noticed in the patient.A transseptal puncture was performed using a medtronic acutus flexcath cross transseptal solution.The caller noted that a smartablate generator was in use, and no bwi ablation catheter was in use.It was noted that the only bwi product in use was the soundstar catheter.The adverse event occurred during the transseptal puncture phase.The patient's blood pressure had dropped to about 49 over 29.The pericardial effusion was confirmed and visualized in the atrium and the ventricle, via ice.Pericardiocentesis was performed, and about 800ml of fluid was removed.The caller reported that the patient is in stable condition.
|
Pericardial effusion is a potential clinical risk associated with the flexcath cross device, which is the same as the risk of a procedure performed with similar, competitive devices.No further information is available as reporter contact information was not provided.If further information is made available a supplemental mdr will be sent.
|