• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2023
Event Type  malfunction  
Event Description
Philips received a complaint from the customer reporting a sensor alarm issue on v60 ventilator, which was later identified as a proximal pressure sensor autozero failed message.The device usage at time of the event is unknown.There was no report of harm, nor patient or user impact.The device did not meet specification for intended use and was removed from service.A philips authorized service provider (asp) evaluated the device and confirmed diagnostic code 110b in the device event log verifying the proximal pressure sensor autozero failed occurrence.The asp replaced gas delivery subsystem (gds) as per service manual.The device was operational and returned to service after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19188687
MDR Text Key341129293
Report Number2518422-2024-22680
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/15/2023
Date Device Manufactured06/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-