Model Number 71992-01 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problems
Fatigue (1849); Hyperglycemia (1905); Nausea (1970); Vomiting (2144); Diabetic Ketoacidosis (2364); Shaking/Tremors (2515); Confusion/ Disorientation (2553)
|
Event Date 04/12/2024 |
Event Type
Injury
|
Event Description
|
An adhesive issue was reported with the adc device.The adc device detached prematurely and the customer was unable to obtain readings and monitor glucose levels.As a result, the customer experienced vomiting, weakness, confusion, tremor, nausea, diabetic ketoacidosis (dka), and a loss of consciousness.The customer was provided rapid insulin (dose unspecified) as treatment by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The dhrs (device history review) for libre sensor kits were reviewed and the dhrs showed the libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An adhesive issue was reported with the adc device.The adc device detached prematurely and the customer was unable to obtain readings and monitor glucose levels.As a result, the customer experienced vomiting, weakness, confusion, tremor, nausea, diabetic ketoacidosis (dka), and a loss of consciousness.The customer was provided rapid insulin (dose unspecified) as treatment by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|
|