COCHLEAR LTD NUCLEUS CI522 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI522
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Pain (1994)
Event Type  malfunction  
Manufacturer Narrative
This report is submitted on april 26, 2024.
 
Event Description
Per the clinic, the patient experienced pain due to magnet dislodgement during an mri (1.5 tesla) on (b)(6) 2024.The patient's head was wrapped as recommended by cochlear.The magnet was manually placed back into correct position without surgical intervention; however the magnet continue to be dislodged.Subsequently, the patient underwent a magnet replacement surgery on (b)(6) 2024.The implanted device remains.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI522 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
asyikin zulkhairi
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key19191571
MDR Text Key341236289
Report Number6000034-2024-01505
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502032599
UDI-Public(01)09321502032599(11)181024(17)201023
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/23/2020
Device Model NumberCI522
Was Device Available for Evaluation? No
Date Manufacturer Received04/03/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-