Catalog Number D134805
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Perforation (2513)
Event Type  Injury  
Manufacturer Narrative
Device investigation details: the device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 31179987l and no internal actions related to the complaint was found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event Description
It was reported that a patient underwent a ventricular tachycardia ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and the patient experienced cardiac perforation that required pericardiocentesis and surgical intervention.During the procedure, the patient's blood pressure started to drop.They checked for effusion and baseline effusion was growing.Pericardiocentesis was performed to drain blood from the pericardial space.Patient wasn't able to be stablized and continued filling.The patient was sent for open heart surgery.Suspected perforation with ablation catheter in the right ventricle (rv) from force (ablation only performed in the left ventricle lv not the right ventricle).High force occurred for breath on anterior free wall of right ventricular outflow tract (rvot) when the ablation catheter entered the rv.Hole.5 x.5 in anterior free wall rvot was found during surgery.Catheters did not seem to fail.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19192124
MDR Text Key341151232
Report Number2029046-2024-01374
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31179987L
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PENTARAY NAV ECO 7FR, D, 2-6-2
Patient Outcome(s) Life Threatening; Required Intervention;
-
-