• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 7122Q/58
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hematoma (1884); Hemorrhage/Bleeding (1888); Unspecified Infection (1930)
Event Date 03/15/2024
Event Type  Injury  
Event Description
It was reported that the patient presented with a large hematoma and pocket infection and bleeding was observed at the implant site.The system was explanted and replaced.A new device was implanted on the right side.The patient was stable.
 
Manufacturer Narrative
Analysis was normal.No anomalies were noted.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The cause of infection could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19193416
MDR Text Key341165746
Report Number2017865-2024-39850
Device Sequence Number1
Product Code NVY
UDI-Device Identifier05414734503174
UDI-Public(01)05414734503174(10)A000152347(17)261231
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Lot NumberA000152347
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/26/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexMale
-
-