GYRUS ACMI, INC. G400 GENERATOR, GYRUS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 777000 |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned and the evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was observed that during the device evaluation, the electrosurgical generator exhibited a faulty auxiliary (aux) board.There were no reports of patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the generator not functioning was due to a faulty auxiliary (aux) and perf (pkrf) board.The cause of the faulty boards could not be determined.Olympus will continue to monitor field performance for this device.
|
|