Model Number 1458Q/86
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Event Description
It was reported that patient presented for a scheduled procedure.During the procedure, it was noted that guidewire could not be advanced through the lead.The lead was not used, and a new lead was implanted.Patient condition was stable.
 
Manufacturer Narrative
The reported event of failure to advance guidewire was confirmed.A complete lead was returned in one piece for analysis.The inner coil was clogged with blood in the distal end region.The cause of the reported event was isolated to the inner coil clogged with blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19197812
MDR Text Key341357520
Report Number2017865-2024-39811
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public(01)05414734503198(10)A000143994(17)260531
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000143994
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/26/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-