Model Number 799-EGB2-00-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "internal comm failure-1471" and "internal communication failure-3470" error messages.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation and the customer report was observed in the forensic memory log.However, the device was put through extensive testing with a known good test setup without duplicating the report.No damage was observed due to the use of the vent in the hyperbaric chamber.The device was recertified and returned to the customer.It is noteworthy to mention; the z vent operator's contains the warning: the device is not intended for hyperbaric operation.Use in a hyperbaric chamber can result in harm to the patient and/or damage to the device.Analysis of reports of this type has not identified an increase in trend.
|
|