• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC BREVERA BREAST BIOPSY SYSTEM; INSTRUMENT, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC BREVERA BREAST BIOPSY SYSTEM; INSTRUMENT, BIOPSY Back to Search Results
Model Number BREVDRV
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2024
Event Type  Injury  
Event Description
It was reported that during a brevera procedure on (b)(6) 2024 the customer reported that the driver device was not working and the needle was inserted into the breast but in the wrong spot.The needle was taken out and put back in but at that moment the needle did not fire.The biopsy could not be completed and the procedure had to be rescheduled.A field engineer examined the equipment and found that the driver had failed.A new driver was installed and the system met the manufacturer´s specifications.No other information is available.
 
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA BREAST BIOPSY SYSTEM
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free park zone b24
san jose 20102
CR   20102
MDR Report Key19201742
MDR Text Key341274577
Report Number1222780-2024-00162
Device Sequence Number1
Product Code KNW
UDI-Device Identifier15420045507760
UDI-Public(01)15420045507760(11)220615
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBREVDRV
Device Catalogue NumberBREVDRV
Was Device Available for Evaluation? No
Date Manufacturer Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-