• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC EVIVA_0913-20; INSTRUMENT, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC EVIVA_0913-20; INSTRUMENT, BIOPSY Back to Search Results
Model Number EVIVA_0913-20
Device Problems Failure to Obtain Sample (2533); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2024
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that during an eviva procedure on (b)(6) 2024 , the needle was inside the breast and the customer could not take the sample.The physician reported that air was leaking from the needle.The biopsy could not be completed due to the physician no longer having functional needles in stock.The patient was rescheduled for a new procedure.No other information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIVA_0913-20
Type of Device
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free park zone b24
san jose 20102
MDR Report Key19202042
MDR Text Key341274473
Report Number1222780-2024-00164
Device Sequence Number1
Product Code KNW
UDI-Device Identifier15420045504974
UDI-Public(01)15420045504974(17)260111(10)E24A12RB
Combination Product (y/n)N
PMA/PMN Number
K180233
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEVIVA_0913-20
Device Catalogue NumberEVIVA_0913-20
Device Lot NumberE24A12RB
Was Device Available for Evaluation? No
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/12/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-