• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number C-VH-3010
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vessel harvesting procedure, t.W.Power supply was not working.They changed power cord, cord to hemopro device, adapter, and checked the switch but power supply did not activate.The power was set to 3.They switched out the power supply and everything worked.There was no reported procedural delay or patient harm.
 
Manufacturer Narrative
E1 event site name exceeded character limitations: (b)(6).Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Manufacturer Narrative
Trackwise #: (b)(4).The reported device is an oem device.The certificate of conformance was reviewed for the serial # (b)(6).The vendor certifies that this device serial # conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key19202121
MDR Text Key341743050
Report Number2242352-2024-00440
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC-VH-3010
Device Catalogue NumberC-VH-3010
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2024
Date Device Manufactured03/11/2021
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
POWER CORD, CORD TO HEMOPRO DEVICE, ADAPTER.
-
-