Catalog Number 500-03-58F
Device Problem Loss of Osseointegration (2408)
Patient Problem Inadequate Osseointegration (2646)
Event Date 08/01/2019
Event Type  Injury  
Manufacturer Narrative
Reported event: an event regarding loosening involving a trident shell was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "there is evidence of loss of fixation of the acetabular component with revision to a pressfit cup with bone graft and mesh in the supra- acetabular region.There is also evidence of stem fracture through the trunnion.Loss of fixation of a pressfit acetabular component and subsequent fracture of the trunnion of the femoral component is confirmed.The root cause of these mechanical failures cannot be determined from the limited documentation provided." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the shell.A review of the provided medical information by a clinical consultant indicated: "there is evidence of loss of fixation of the acetabular component with revision to a pressfit cup with bone graft and mesh in the supra- acetabular region.There is also evidence of stem fracture through the trunnion.Loss of fixation of a pressfit acetabular component and subsequent fracture of the trunnion of the femoral component is confirmed.The root cause of these mechanical failures cannot be determined from the limited documentation provided." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported the patient had a hip cup revision due to aseptic loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRIT HEMI SOLIDBACK CUP 58MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key19202138
MDR Text Key341272048
Report Number0002249697-2024-00659
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327040050
UDI-Public07613327040050
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K081171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Catalogue Number500-03-58F
Device Lot NumberMJKN2M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexMale
Patient Weight150 KG
-
-