Sakura finetek usa was notified by sakura finetek europe on april 25,2024 that a local service engineer visited the customer on 19-mar-2024 to collect the instruments logfiles, reagent samples and to check and clean the instrument.An obstruction was found in the connector of station number 4 of the instrument which was then removed.After this, the customer did not report any issues with the use of the instrument.Logfiles were analysed by technical support and found no technical defects in vacuum, pressure, temperatures.Instrument performed according to technical specifications.Logfiles and reagents samples of the run in question were analysed by application support.It was found that the customer isn't performing the weekly warm water flushes according to the recommended cleaning procedures as stated by the manufacturer in the operating manual of the device.Not performing warm water flushes on a regular base, increases the of buildup of formalin crystals and eventually blockage of tubing.This is the most likely rootcause of the issue that the customer experienced.Reagent samples were measured and found to be of the correct values, nothing out of specification was observed.Sakura finetek europe has not received any information on patient impact by the time of this report.
|
Sakura finetek europe notified sakuta finetek usa on 28-mar-2024 about the incident that occured on (b)(6) 2024, the customer experienced an issue with the tissue-tek vip 6 ai on 12-mar-2024 where the instrument stopped processing at station 4 (alcohol).They had contacted a sakura employee who talked them through the issue and they were able completed the run without impact on tissue.After this run, the customer performed a cleaning with water and acetic acid.A new run was initiated and started in the night from 13th on 14th of march without testing the instrument and the issue reoccurred.This time, the samples were hard to cut and impact on patient has not yet been communicated to sakura finetek europe.
|