Model Number VH-4000
Device Problem Device Remains Activated (1525)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vessel harvesting procedure after vasoview hemopro 2 was activated, it continued to activate, including the sound, after the toggle had been released.The device was removed from the cannula, the jaws were opened, and the beep could still be heard.It didn¿t appear to still be generating any heat at that point.Harvester unplugged the cord and removed it from the field.Another device was opened and used with the same cord with no further issues.There was a slight procedural delay while a new device was being opened.There was no adverse event or other issues.
 
Event Description
N/a.
 
Manufacturer Narrative
Trackwise #: (b)(4).Updated sections: b-4, d-4 (lot and exp date), d-9,g-3, g-6, h-2,h-3,h-4, h-6, h-10 h3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key19202945
MDR Text Key341299047
Report Number2242352-2024-00442
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000332922
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/09/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
-
-