It was reported that the patient was admitted to the hospital due to general unwellness.A low blood pressure was detected and was taken care of clinically.Looking at the log file history, the power reading was noted to increase from the usual average of 5 watts to over 9 watts.The fixed speed was not able to be reached despite well filled left ventricle.The log files were downloaded and analyzed and high readings in power, rotor noise, and temperature were detected.The high power consumption was reported to be due to the imbalance of the rotor which caused the motor to work harder to maintain levitation.The scenario was said to has probably occurred due to an unstable rotor and some form of occlusion.The only confirmation for the occlusion were reported to be the readings of the pump parameters.The patient underwent an echocardiogram.However, the images could not be provided.Other than the low blood pressure the patient had slight nausea and loose stool.Their medication was adjusted to increase the blood pressure slightly.The patient was stable.
|