MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4483
Device Problems Failure to Disconnect (2541); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  malfunction  
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the twist clamp connector of a minicap transfer set ¿unscrewed from the transfer set¿ and the transfer set would not come apart from the patient line of the cassette causing the home patitient (hp) to cut the line.This was identified while attempting to disconnect from peritoneal dialysis (pd) therapy.The transfer set was replaced.There was no report of patient injury, however, the hp received prophylactic treatment as a preventative measure.The patient did not develop any symptoms.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19203562
MDR Text Key341298372
Report Number1416980-2024-01948
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4483
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AMIA CASSETTE; AMIA DEVICE
Patient SexFemale
-
-