Model Number G247
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Shock from Patient Lead(s) (3162)
Event Date 02/11/2024
Event Type  Injury  
Event Description
The health care professional (hcp) requested technical services (ts) review of device data to confirm device operation.It was noted that the patient with this cardiac resynchronization therapy defibrillator (crt-d) received inappropriate anti-tachycardia pacing (atp) and inappropriate shocks.It was noted pacemaker mediated tachycardia (pmt) episodes were also recorded.Upon review of the presenting electrogram (egm) showed that the inappropriate therapy and pmt episodes appeared to be due to atrial fibrillation (af) with rapid ventricular response (rvr).It was noted that the shock therapy had difficulties converting the rhythm.At this time, the crt-d remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19203703
MDR Text Key341293115
Report Number2124215-2024-25783
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/25/2024
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number278674
Was Device Available for Evaluation? No
Date Manufacturer Received04/06/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age56 YR
Patient SexMale
-
-