Model Number CYF-VHA
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2024
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the cyto-nephro videoscope was not correctly reprocessed, the water supply pipeline of the reprocessor had never been disinfected, and the water filter had never been changed after delivery to the user facility.The issue was found during reprocessing.There were no reports of patient involvement.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over seven (7) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined as the device was not returned for evaluation.Although, the customer confirmed that the device was incorrectly reprocessed it cannot be specified as to why.The event can be detected/prevented by handling the device in accordance with the following instructions for use (ifu): instruction manual endoscope cleaning and disinfection equipment oer-s(acecide 6% for disinfectant)operation edition; chapter 3 inspection before use; chapter 4 basic procedure of cleaning and disinfecting endoscopes; chapter 6 other functions.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19203917
MDR Text Key341933157
Report Number3002808148-2024-03982
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170411274
UDI-Public04953170411274
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OER-S-ENDOSCOPE REPROCESSOR
-
-