• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G247
Device Problems Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2024
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) stored pacemaker mediated tachycardia (pmt) episodes.Upon further review of the episodes, boston scientific technical services (ts) recommended further evaluation of the right atrial (ra) lead due to signals following the atrial pace that could be either far field oversensing or loss of capture (loc).The involved lead is a non boston scientific product.A chest x ray was recommended.No adverse patient effects were reported.At this time, this device system remains in service.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) stored pacemaker mediated tachycardia (pmt) episodes.Upon further review of the episodes, boston scientific technical services (ts) recommended further evaluation of the right atrial (ra) lead due to signals following the atrial pace that could be either far field oversensing or loss of capture (loc).The involved lead is a non boston scientific product.A chest x ray was recommended.Further information was provided that a chest x ray was performed to evaluate atrial lead placement and it was confirmed to be in a good position.It was determined that this patient had a junctional escape rhythm.The physician increased the lower rate from 50 to 60 beats per minute (bpm) and will continue to monitor the patient.No adverse patient effects were reported.At this time, this device system remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19205665
MDR Text Key341486171
Report Number2124215-2024-25810
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number315561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-