It was reported that this cardiac resynchronization therapy defibrillator (crt-d) stored pacemaker mediated tachycardia (pmt) episodes.Upon further review of the episodes, boston scientific technical services (ts) recommended further evaluation of the right atrial (ra) lead due to signals following the atrial pace that could be either far field oversensing or loss of capture (loc).The involved lead is a non boston scientific product.A chest x ray was recommended.No adverse patient effects were reported.At this time, this device system remains in service.
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) stored pacemaker mediated tachycardia (pmt) episodes.Upon further review of the episodes, boston scientific technical services (ts) recommended further evaluation of the right atrial (ra) lead due to signals following the atrial pace that could be either far field oversensing or loss of capture (loc).The involved lead is a non boston scientific product.A chest x ray was recommended.Further information was provided that a chest x ray was performed to evaluate atrial lead placement and it was confirmed to be in a good position.It was determined that this patient had a junctional escape rhythm.The physician increased the lower rate from 50 to 60 beats per minute (bpm) and will continue to monitor the patient.No adverse patient effects were reported.At this time, this device system remains in service.
|