Updated field: b4, d9, g3, g6, h2, h3, h4, h6(type of investigation, investigation findings, investigation conclusions),h10.Additional contcat details: event site postal code: (b)(6).It was reported that the cardiosave intra-aortic balloon pump (iabp) power cord cannot be extended.There was no patient involvement and no patient harm reported.A getinge field service engineer fse was dispatched to the site to evaluate the unit.He states that the power cord of the machine was stuck and could not be extended.The machine was opened on site and the external power cord of the winder was reorganized to restore normal operation.It is recommended that users keep a constant speed when retracting the power cord.The device was returned to customer for clinical use.
|